Cronologic Ndigo Crate Guide de l'utilisateur Page 10

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 57
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 9
+
-
450MHz
offset
Figure 2.4: Input circuit for each of the four analog channels.
2.2.2 Analog Inputs
The analog inputs of the ADC are single ended LEMO00 coax connectors. The inputs have a
50Ω impedance and are DC coupled. The inputs are converted to a differential signal using an
operational amplifier. The input voltage range is 1V
p-p
centered around a programmable offset
voltage. See section 6.4.3 for details.
Analog Offsets
+0.5V
analog_offset[i] = 0V
-0.5 V
+0.5 V
analog_offset[i] = 0.2V
-0.5 V
0.2V
Figure 2.5: Users can add analog offset to the input before sampling
Users can set the approximate center of the input voltage range individually for each channel
by using the analog offset parameter. The valid voltage range for the offset is specified in section
cronologic GmbH & Co. KG 6 Ndigo250M-14 User Guide